Product details
Product details Lunaire steady-state stability chambers are available in three sizes and feature a large workspace with multiple temperatures and humidity combinations. These steady-state stability chambers meet a variety of testing needs and are designed for quick shipping and easy installation. The temperature and humidity settings are housed in a single controller which allows for hands-free operation. High Low Limit Alarms High Capacity (10, 20, and 30 cubic foot workspace) High Performance -20°C to 200°C temperature range ±0.3° after stabilization 20% to 95% RH in the dry bulb range of +20°C to +85°C, limited by a 5°C dewpoint ±2% for relative humidity ICH Q1A Stability Testing Long Term Performance: 25°C/60% RH or 30°C/60% RH Intermediate Performance: 30°C/65% RH Accelerated Performance: 40°C/75% RH Highly Energy Efficient Set up and monitoring is fast and easy Learn About Reverse Osmosis Options